Pulse rate (PR)
Average 75 bpm
FROM SENSING TO ANALYSIS
Optical sensing captures fingertip photoplethysmography (PPG) signals. Signal quality control and AI waveform analysis then estimate the arterial blood pressure waveform, showing pressure changes throughout the cardiac cycle.
Optical sensing
Red and infrared light capture fingertip pulse waves. A stable optical sensor and ambient-light rejection reduce interference from external light sources, preserving waveform detail and variation.
Signal processing
On-device filtering and signal-quality assessment reduce noise and select pulse signals that meet quality thresholds, providing reliable input for subsequent AI analysis.
AI analysis
Personalized deep learning models estimate arterial blood pressure waveforms from optical pulse signals and derive 32 cardiovascular indicators, revealing physiological information beyond systolic and diastolic pressure.
CLINICAL VALIDATION
Measurements from 111 participants at Hualien Tzu Chi Hospital were compared with a mercury sphygmomanometer to assess blood pressure measurement performance against ISO 81060-2:2018.
| Assessment criterion | Threshold | Systolic | Diastolic | Result |
|---|---|---|---|---|
| Mean errormmHg | ±5 | −2.10 | 2.44 | Met |
| Standard deviation of errormmHg | ≤8 | 6.20 | 5.82 | Met |
| Standard deviation of paired differencesmmHg | ≤6.62 / 6.51 | 6.60 | 6.35 | Met |
RECORDS OVER TIME
Review changes across different periods with trend charts, making each measurement part of a longer-term picture of health and care.
Average 75 bpm
Average 97 %
Average 121 / 74 mmHg
Average 48 ms